Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device

Sarah Gruber, Angela Nickel

Frontiers in Medical Technology, 5, 1195529 · Published 14 June 2023 · DOI 10.3389/fmedt.2023.1195529

Open PDFView at the publisherCite this paper

In brief

“Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device” is a 2023 journal article by Sarah Gruber and Angela Nickel, published in Frontiers in Medical Technology. OpenAlex counts 257 citations as of 25 November 2026, which puts it in the top 8% of works published in 2023 by citations. It is open access under a CC BY licence: the full text is free to read at Frontiers in Medical Technology.

Citations
257
Most cited in
2026 (159)
Field-weighted impact
46.37
References
29
Authors
2

Abstract

Background: cytotoxicity tests. This test evaluates the effects of medical device use on cell health. The existence of the specific standard suggests that the tests will produce reliable and comparable results. However, the ISO 10993-5 offers wide latitude in the test specifications. In the past, we noticed inconsistencies of the results from different laboratories. Objective: To determine if the specifications of the standard ISO 10993-5 are explicit to ensure the comparability of test results and, if not, identify potential influencing factors. Methods: cytotoxicity test according to ISO 10993-5. Fifty-two international laboratories evaluated the cytotoxicity for two unknown samples. One was polyethylene (PE) tubing, which is expected to be non-cytotoxic and the other was polyvinyl chloride (PVC) tubing, for which a cytotoxic potential was presumed. All laboratories were asked to perform an elution test with predefined extraction specifications. The other test parameters were freely chosen by the laboratories according to the guidelines set by the standard. Results: To our surprise only 58 percent of the participating laboratories identified the cytotoxic potential of both materials as expected. Particularly for PVC a considerable variation of the results between the laboratories was observed [mean = 43 ± 30 (SD), min = 0, max = 100]. We showed that ten percent serum supplementation to the extraction medium, as well as longer incubation of the cells with the extract, greatly increased the test sensitivity for PVC. Conclusion: The results clearly show that the specifications set by the ISO 10993-5 are not explicit enough to obtain comparable results for an identical medical device. To set requirements that ensure reliable cytotoxicity assessments, further research will be necessary to identify the best test conditions for specific materials and/or devices and the standard needs to be revised accordingly.

Cite this paper

Formatted from the catalogue record. Check names, capitals and page numbers against the paper before you submit.

APA 7

Gruber, S., & Nickel, A. (2023). Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device. Frontiers in Medical Technology, 5, 1195529. https://doi.org/10.3389/fmedt.2023.1195529

MLA 9

Gruber, Sarah, and Angela Nickel. “Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device.” Frontiers in Medical Technology, vol. 5, 2023, p. 1195529. https://doi.org/10.3389/fmedt.2023.1195529.

Chicago

Gruber, Sarah and Angela Nickel. 2023. “Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device.” Frontiers in Medical Technology 5: 1195529. https://doi.org/10.3389/fmedt.2023.1195529.

Harvard

Gruber, S. and Nickel, A. (2023) ‘Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device’, Frontiers in Medical Technology, 5, p. 1195529. doi:10.3389/fmedt.2023.1195529.

IEEE

S. Gruber and A. Nickel, “Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device,” Frontiers in Medical Technology, vol. 5, p. 1195529, 2023, doi: 10.3389/fmedt.2023.1195529.

BibTeX

@article{gruber2023toxic,
  title     = {{Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device}},
  author    = {Gruber, Sarah and Nickel, Angela},
  journal   = {Frontiers in Medical Technology},
  year      = {2023},
  volume    = {5},
  pages     = {1195529},
  doi       = {10.3389/fmedt.2023.1195529},
  url       = {https://doi.org/10.3389/fmedt.2023.1195529}
}

Download for a reference manager: BibTeX (.bib) · RIS for Zotero, Mendeley and EndNote (.ris)

Citations by year

YearCitationsShare of the best year
202322
20242424
20257272
2026159159

Its field-weighted citation impact is 46.37: it has been cited 46.37 times as often as the average work of the same field, type and year (1.00 is the world average).

Topic and keywords

Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Healthcare Regulation

  • ISO 10993-5
  • cytotoxicity assay
  • medical device safety
  • inter-laboratory variability
  • test result comparability
  • PVC pipes
  • polyethylene pipes
  • elution test

Work with this paper

Most-cited papers in Medicine

All highly cited Medicine papers

Questions about this paper

How do I cite this paper in APA style?

Gruber, S., & Nickel, A. (2023). Toxic or not toxic? The specifications of the standard ISO 10993-5 are not explicit enough to yield comparable results in the cytotoxicity assessment of an identical medical device. Frontiers in Medical Technology, 5, 1195529. https://doi.org/10.3389/fmedt.2023.1195529

How many times has this paper been cited?

257 times, according to OpenAlex on 25 November 2026. Google Scholar and Scopus count different sets of sources, so their numbers differ.

Is this paper free to read?

Yes. It is open access at Frontiers in Medical Technology: sign in and open the PDF from this page.

Who wrote this paper?

Sarah Gruber, Angela Nickel.

What is the DOI of this paper?

10.3389/fmedt.2023.1195529, permanently at https://doi.org/10.3389/fmedt.2023.1195529. Sign in and open it from this page.

Bibliographic data from OpenAlex (CC0), last checked 25 November 2026. ResearchGuru does not host this paper; once you are signed in, the links open it at the publisher or the repository that does.