“Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video” is a 2023 journal article by Benjamin K Miller and 2 co-authors, published in Microscopy and Microanalysis. OpenAlex counts 0 citations as of 16 November 2026. It is open access: the full text is free to read at Microscopy and Microanalysis.
- Citations
- 0
- References
- 2
- Authors
- 3
Abstract
While transmission electron microscopy (TEM) can elucidate the structure of nanomaterials from the micro to the atomic scale, many functional materials are beam sensitive. This has made TEM characterization difficult or even impossible for some materials. The definition of “beam-sensitive” can vary greatly depending on who defines it and at what resolution the sample is observed. For the single-particle cryo-electron microscopy (cryo-EM) community, an accumulated dose of just tens of electrons per angstrom squared (e-/Å2) is enough to destroy a sample, but atomic resolution imaging of an individual protein is not expected [1,2]. For researchers studying materials at the atomic scale, a total dose of several thousand e-/Å2 may be considered low-dose, especially if the dose rate is low [3]. In contrast, for traditional atomic resolution imaging of robust TEM samples, the total dose may be on the order of 104-105 e-/Å2. Techniques that capture more than one image, like tomography, 4D STEM, or in-situ experiments, also tend to require larger electron doses. Whatever the magnitude of the critical dose above which a structure is likely to be significantly damaged, observation of the pristine material without damage is desired. Recently developed counting direct detection cameras can make observation of the pristine structure possible. Cameras like Gatan’s K2® and K3®, recent ThermoFisher Scientific Falcon cameras, and the Direct Electron Apollo can count each electron individually, significantly reducing readout noise and making it possible to operate the microscope at a very low dose rate on the sample. The exact range varies by camera and by the TEM magnification. In 2022 Gatan released two new counting cameras, the Alpine™, and the Metro™, optimized for low-dose, low-kV data collection. While the Alpine is specifically designed for cryo-EM applications, the Metro is intended for materials science research and exhibits excellent performance, whether capturing imaging, diffraction, or in-situ data. To prove that the pristine structure has been observed for a given material, one should observe the sample at a dose rate at which damage is expected to be slow and monitor the structural changes which are observed as a function of time. Ideally, these changes should be quantified. Since the Metro camera has built-in in-situ capability, continuous videos can be captured. This, combined with the drift correction capabilities of DigitalMicrograph®, makes observation and quantification of structural changes much easier. We can also observe the sample at a very low dose rate over a long period of time or even change the dose rate during a single video acquisition to quantify the damage accumulation. In this work…
Cite this paper
Formatted from the catalogue record. Check names, capitals and page numbers against the paper before you submit.
APA 7
Miller, B. K., Xu, M., & Czarnik, C. (2023). Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video. Microscopy and Microanalysis, 29(Supplement_1), 1835–1837. https://doi.org/10.1093/micmic/ozad067.950
MLA 9
Miller, Benjamin K, et al. “Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video.” Microscopy and Microanalysis, vol. 29, no. Supplement_1, 2023, pp. 1835–1837. https://doi.org/10.1093/micmic/ozad067.950.
Chicago
Miller, Benjamin K, Mingjie Xu, and Cory Czarnik. 2023. “Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video.” Microscopy and Microanalysis 29 (Supplement_1): 1835–1837. https://doi.org/10.1093/micmic/ozad067.950.
Harvard
Miller, B. K., Xu, M. and Czarnik, C. (2023) ‘Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video’, Microscopy and Microanalysis, 29(Supplement_1), pp. 1835–1837. doi:10.1093/micmic/ozad067.950.
IEEE
B. K. Miller, M. Xu, and C. Czarnik, “Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video,” Microscopy and Microanalysis, vol. 29, no. Supplement_1, pp. 1835–1837, 2023, doi: 10.1093/micmic/ozad067.950.
BibTeX
@article{miller2023assessing,
title = {{Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video}},
author = {Miller, Benjamin K and Xu, Mingjie and Czarnik, Cory},
journal = {Microscopy and Microanalysis},
year = {2023},
volume = {29},
number = {Supplement_1},
pages = {1835--1837},
doi = {10.1093/micmic/ozad067.950},
url = {https://doi.org/10.1093/micmic/ozad067.950}
}
Download for a reference manager: BibTeX (.bib) · RIS for Zotero, Mendeley and EndNote (.ris)
Topic and keywords
Life Sciences › Biochemistry, Genetics and Molecular Biology › Structural Biology › Advanced Electron Microscopy Techniques and Applications
- critical dose
- beam-sensitive materials
- electron dose rate
- structural damage
- in situ electron diffraction
Work with this paper
Most-cited papers in Biochemistry, Genetics and Molecular Biology
- PROTEIN MEASUREMENT WITH THE FOLIN PHENOL REAGENTOliver H. Lowry, NiraJ. Rosebrough, A. Lewis Farr and 1 more · Journal of Biological Chemistry · 1951 · 316,470 citations
- A Rapid and Sensitive Method for the Quantitation of Microgram Quantities of Protein Utilizing the Principle of Protein-Dye BindingM Bradford · Analytical Biochemistry · 1976 · 256,780 citations
- Analysis of Relative Gene Expression Data Using Real-Time Quantitative PCR and the 2−ΔΔCT MethodKenneth J. Livak, Thomas D. Schmittgen · Methods · 2001 · 184,418 citations
- Molecular Cloning: A Laboratory ManualJoseph Sambrook, Elisabeth Fritsch, Tom Maniatis · UCLA - Biblioteca de Ciencias de la Salud · 1989 · 133,493 citations
- Basic Local Alignment Search ToolStephen F. Altschul, Warren R. Gish, Webb Miller and 2 more · Journal of Molecular Biology · 1990 · 100,529 citations
All highly cited Biochemistry, Genetics and Molecular Biology papers
Questions about this paper
How do I cite this paper in APA style?
Miller, B. K., Xu, M., & Czarnik, C. (2023). Assessing Critical Dose for Beam-sensitive Samples using Low-dose Counted in-situ Video. Microscopy and Microanalysis, 29(Supplement_1), 1835–1837. https://doi.org/10.1093/micmic/ozad067.950
How many times has this paper been cited?
0 times, according to OpenAlex on 16 November 2026. Google Scholar and Scopus count different sets of sources, so their numbers differ.
Is this paper free to read?
Yes. It is open access: read the free full text at Microscopy and Microanalysis.
Who wrote this paper?
Benjamin K Miller, Mingjie Xu, Cory Czarnik.
What is the DOI of this paper?
10.1093/micmic/ozad067.950. The permanent link is https://doi.org/10.1093/micmic/ozad067.950.
Bibliographic data from OpenAlex (CC0), last checked 16 November 2026. ResearchGuru does not host this paper; the links go to the publisher or the repository that does.