Journal article · 2008

Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method

Vratislav Chromý, Lenka Šváchová, Lukáš Novosád, Jiří Jarkovský, Pavel Sedlák, Petr Horák, Hana Dobrovolná, Barbora Hlaváčová

Clinical Chemistry and Laboratory Medicine (CCLM), 47(1), 91–101 · Published 31 December 2008 · DOI 10.1515/cclm.2009.011

View at the publisherCite this paper

In brief

“Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method” is a 2008 journal article by Vratislav Chromý and 7 co-authors, published in Clinical Chemistry and Laboratory Medicine (CCLM). OpenAlex counts 15 citations as of 15 November 2026. The publisher's version is not open access, and OpenAlex records no free legal copy.

Citations
15
Most cited in
2014 (3)
Field-weighted impact
0.74
References
15
Authors
8

Abstract

BACKGROUND: Assay of total serum protein by the biuret method calibrated with albumin standards according to the reference method provides results with a positive bias approximately 3%-5% exceeding the total error of 3.4% allowable for total protein in serum analysis made by analysers using two-part reagents and short-term procedures. METHODS: We used two types of… Read the full abstract at the publisher.

Excerpt. This paper is not open access.

Cite this paper

Formatted from the catalogue record. Check names, capitals and page numbers against the paper before you submit.

APA 7

Chromý, V., Šváchová, L., Novosád, L., Jarkovský, J., Sedlák, P., Horák, P., Dobrovolná, H., & Hlaváčová, B. (2008). Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method. Clinical Chemistry and Laboratory Medicine (CCLM), 47(1), 91–101. https://doi.org/10.1515/cclm.2009.011

MLA 9

Chromý, Vratislav, et al. “Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method.” Clinical Chemistry and Laboratory Medicine (CCLM), vol. 47, no. 1, 2008, pp. 91–101. https://doi.org/10.1515/cclm.2009.011.

Chicago

Chromý, Vratislav, Lenka Šváchová, Lukáš Novosád, Jiří Jarkovský, Pavel Sedlák, Petr Horák, Hana Dobrovolná, and Barbora Hlaváčová. 2008. “Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method.” Clinical Chemistry and Laboratory Medicine (CCLM) 47 (1): 91–101. https://doi.org/10.1515/cclm.2009.011.

Harvard

Chromý, V. et al. (2008) ‘Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method’, Clinical Chemistry and Laboratory Medicine (CCLM), 47(1), pp. 91–101. doi:10.1515/cclm.2009.011.

IEEE

V. Chromý et al., “Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method,” Clinical Chemistry and Laboratory Medicine (CCLM), vol. 47, no. 1, pp. 91–101, 2008, doi: 10.1515/cclm.2009.011.

BibTeX

@article{chromy2008albumin,
  title     = {{Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method}},
  author    = {Chromý, Vratislav and Šváchová, Lenka and Novosád, Lukáš and Jarkovský, Jiří and Sedlák, Pavel and Horák, Petr and Dobrovolná, Hana and Hlaváčová, Barbora},
  journal   = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year      = {2008},
  volume    = {47},
  number    = {1},
  pages     = {91--101},
  doi       = {10.1515/cclm.2009.011},
  url       = {https://doi.org/10.1515/cclm.2009.011}
}

Download for a reference manager: BibTeX (.bib) · RIS for Zotero, Mendeley and EndNote (.ris)

Citations by year

YearCitationsShare of the best year
201211
201322
201433
201533
201722
201911
202222

Its field-weighted citation impact is 0.74: it has been cited 0.74 times as often as the average work of the same field, type and year (1.00 is the world average).

Topic and keywords

Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control

  • serum total protein
  • biuret method
  • positive bias
  • assay interference
  • bilirubin
  • Kjeldahl method
  • automated analyzers
  • reference method comparison

Work with this paper

Most-cited papers in Health Professions

All highly cited Health Professions papers

Questions about this paper

How do I cite this paper in APA style?

Chromý, V., Šváchová, L., Novosád, L., Jarkovský, J., Sedlák, P., Horák, P., Dobrovolná, H., & Hlaváčová, B. (2008). Albumin-based or albumin-linked calibrators cause a positive bias in serum proteins assayed by the biuret method. Clinical Chemistry and Laboratory Medicine (CCLM), 47(1), 91–101. https://doi.org/10.1515/cclm.2009.011

How many times has this paper been cited?

15 times, according to OpenAlex on 15 November 2026. Google Scholar and Scopus count different sets of sources, so their numbers differ.

Is this paper free to read?

OpenAlex records no free legal copy. Check your library's access, the authors' pages or a repository, or ask the corresponding author for a copy.

Who wrote this paper?

Vratislav Chromý, Lenka Šváchová, Lukáš Novosád, Jiří Jarkovský, Pavel Sedlák, Petr Horák, Hana Dobrovolná, Barbora Hlaváčová.

What is the DOI of this paper?

10.1515/cclm.2009.011. The permanent link is https://doi.org/10.1515/cclm.2009.011.

Bibliographic data from OpenAlex (CC0), last checked 15 November 2026. ResearchGuru does not host this paper; the links go to the publisher or the repository that does.