Journal article · 2023 · Open access
Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study
Journal of Prosthodontic Research, 68(1), 172–180 · Published 1 January 2023 · DOI 10.2186/jpr.jpr_d_22_00251
In brief
“Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study” is a 2023 journal article by Alessandro Pozzi and 5 co-authors, published in Journal of Prosthodontic Research. OpenAlex counts 51 citations as of 24 November 2026, which puts it in the top 8% of works published in 2023 by citations. It is open access under a CC BY-NC licence: the full text is free to read at Journal of Prosthodontic Research.
- Citations
- 51
- Most cited in
- 2026 (23)
- Field-weighted impact
- 9.16
- References
- 26
- Authors
- 6
Abstract
PURPOSE: To assess and compare the accuracies of intraoral scanners (IOS) and stereophotogrammetry (SPG) devices for complete-arch digital implant impressions. METHODS: A 4-analog model was digitized using a desk scanner to obtain a reference file. Thirty test scans were conducted using the investigated IOS device, while an additional 30 scans were performed using the SPG device. Using the best-fit algorithm, the resulting 60 test files were aligned with the reference file. Linear (ΔX, ΔY, and ΔZ-axis) and angular deviations (ΔANGLE) were evaluated. Three-dimensional (3D) deviation was calculated based on the Euclidean distance (ΔEUC). The analysis was stratified according to the scanning device and implant position. Fisher's F and t-tests were used to compare the variances and expected values of the two scanning systems. RESULTS: IOS expressed a higher 3D (ΔEUC) mean deviation than SPG (52.8 µm vs. 33.4 µm, P < 0.0001), with extreme measurements up to 181.9 µm. A significantly higher standard deviation (SD) was associated with IOS (37.1 µm vs. 17.7 µm, P < 0.0001). Considering angular deviations, the IOS showed slightly higher angular mean deviations (ΔANGLE) than the SPG (0.28° vs. 0.24°, P = 0.0022), with extreme measurements of up to 0.73°. The SPG SD values were significantly lower than the IOS SD values (0.14° vs. 0.04°, P < 0.0001). CONCLUSIONS: The SPG showed significantly higher 3D and angular accuracies for complete arch implant impressions, with consistent repeatability. IOS scanning revealed significantly higher extreme deviations exceeding the acceptable threshold value. Despite study limitations, SPG appears more feasible than IOS for complete-arch digital implant impressions.
Cite this paper
Formatted from the catalogue record. Check names, capitals and page numbers against the paper before you submit.
APA 7
Pozzi, A., Agliardi, E., Lio, F., Nagy, K., Nardi, A., & Arcuri, L. (2023). Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study. Journal of Prosthodontic Research, 68(1), 172–180. https://doi.org/10.2186/jpr.jpr_d_22_00251
MLA 9
Pozzi, Alessandro, et al. “Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study.” Journal of Prosthodontic Research, vol. 68, no. 1, 2023, pp. 172–180. https://doi.org/10.2186/jpr.jpr_d_22_00251.
Chicago
Pozzi, Alessandro, Enrico Agliardi, Fabrizio Lio, Katalin Nagy, Alessandra Nardi, and Lorenzo Arcuri. 2023. “Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study.” Journal of Prosthodontic Research 68 (1): 172–180. https://doi.org/10.2186/jpr.jpr_d_22_00251.
Harvard
Pozzi, A. et al. (2023) ‘Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study’, Journal of Prosthodontic Research, 68(1), pp. 172–180. doi:10.2186/jpr.jpr_d_22_00251.
IEEE
A. Pozzi, E. Agliardi, F. Lio, K. Nagy, A. Nardi, and L. Arcuri, “Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study,” Journal of Prosthodontic Research, vol. 68, no. 1, pp. 172–180, 2023, doi: 10.2186/jpr.jpr_d_22_00251.
BibTeX
@article{pozzi2023accuracy,
title = {{Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study}},
author = {Pozzi, Alessandro and Agliardi, Enrico and Lio, Fabrizio and Nagy, Katalin and Nardi, Alessandra and Arcuri, Lorenzo},
journal = {Journal of Prosthodontic Research},
year = {2023},
volume = {68},
number = {1},
pages = {172--180},
doi = {10.2186/jpr.jpr_d_22_00251},
url = {https://doi.org/10.2186/jpr.jpr_d_22_00251}
}
Download for a reference manager: BibTeX (.bib) · RIS for Zotero, Mendeley and EndNote (.ris)
Citations by year
| Year | Citations | Share of the best year |
|---|---|---|
| 2023 | 2 | |
| 2024 | 4 | |
| 2025 | 22 | |
| 2026 | 23 |
Its field-weighted citation impact is 9.16: it has been cited 9.16 times as often as the average work of the same field, type and year (1.00 is the world average).
Topic and keywords
Health Sciences › Dentistry › Oral Surgery › Dental Implant Techniques and Outcomes
- intraoral scanners
- stereophotogrammetry
- digital implant impressions
- scanning accuracy
- 3D deviation
- angular deviation
- dental implants
- digital dentistry
Work with this paper
Most-cited papers in Dentistry
- Periodontal Disease in Pregnancy II. Correlation Between Oral Hygiene and Periodontal ConditionJohn Silness, Harald A. Löe · Acta Odontologica Scandinavica · 1964 · 8,071 citations
- Periodontal Disease in Pregnancy I. Prevalence and SeverityHarald A. Löe, John Silness · Acta Odontologica Scandinavica · 1963 · 7,339 citations
- Microbial complexes in subgingival plaqueS. S. Socransky, Anne D. Haffajee, MaryAnn Cugini and 2 more · Journal Of Clinical Periodontology · 1998 · 5,207 citations
- Development of a Classification System for Periodontal Diseases and ConditionsGary C. Armitage · Annals of Periodontology · 1999 · 5,198 citations
- Diagnostic Criteria for Temporomandibular Disorders (DC/TMD) for Clinical and Research Applications: Recommendations of the International RDC/TMD Consortium Network* and Orofacial Pain Special Interest Group†ERIC L. SCHIFFMAN, Richard K. Ohrbach, Edmond L. Truelove and 27 more · Journal of Oral & Facial Pain and Headache · 2014 · 4,942 citations
Questions about this paper
How do I cite this paper in APA style?
Pozzi, A., Agliardi, E., Lio, F., Nagy, K., Nardi, A., & Arcuri, L. (2023). Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study. Journal of Prosthodontic Research, 68(1), 172–180. https://doi.org/10.2186/jpr.jpr_d_22_00251
How many times has this paper been cited?
51 times, according to OpenAlex on 24 November 2026. Google Scholar and Scopus count different sets of sources, so their numbers differ.
Is this paper free to read?
Yes. It is open access at Journal of Prosthodontic Research: sign in and open the PDF from this page.
Who wrote this paper?
Alessandro Pozzi, Enrico Agliardi, Fabrizio Lio, Katalin Nagy, Alessandra Nardi, Lorenzo Arcuri.
What is the DOI of this paper?
10.2186/jpr.jpr_d_22_00251, permanently at https://doi.org/10.2186/jpr.jpr_d_22_00251. Sign in and open it from this page.
Bibliographic data from OpenAlex (CC0), last checked 24 November 2026. ResearchGuru does not host this paper; once you are signed in, the links open it at the publisher or the repository that does.