Journal article · 2025 · Open access
Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study
Journal of Dentistry, 164, 106254 · Published 20 November 2025 · DOI 10.1016/j.jdent.2025.106254
In brief
“Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study” is a 2025 journal article by Henriette Lerner and 5 co-authors, published in Journal of Dentistry. OpenAlex counts 9 citations as of 17 November 2026, which puts it in the top 2% of works published in 2025 by citations. It is open access under a CC BY licence: the full text is free to read at Journal of Dentistry.
- Citations
- 9
- Most cited in
- 2026 (9)
- Field-weighted impact
- 6.38
- References
- 53
- Authors
- 6
Abstract
OBJECTIVE: To evaluate the accuracy of full-arch (FA) implant scan with a novel system (Nexus IOS®, Osteon Medical) using scan gauges (SGs), and to compare it with that of different conventional scan body (SB) systems that are currently available. METHODS: A plaster model with 4 multi-unit-abutment analogs was prepared and inter-implant distances between its analogs were measured using a coordinate measuring machine (CMM), to obtain reference distances. An expert operator used an intraoral scanner (IOS) to capture scans with five different systems (Nexus IOS®, DESS®, Medentika®, IPD® and Elos®, n = 10 scans per group). These measurements were compared with the reference CMM-generated model to calculate inter-implant distance trueness. The study design addressed three questions: 1) Are there overall differences among the systems? 2) Are there overall differences among segments/points? and 3) For a given segment did systems perform differently? A statistical analysis was performed. RESULTS: Among the systems, Nexus IOS® system exhibited the highest trueness, followed by DESS®. Medentika® and IPD® showed slightly higher deviations and variability, indicating moderate performance. Elos® had the least trueness among the systems tested, with the highest median deviation and the widest overall spread, suggesting greater inconsistency in dimensional accuracy. CONCLUSIONS: Within the limitations of this in vitro study, the Nexus IOS® SGs demonstrated the highest dimensional accuracy, showing superior trueness and precision in FA implant scanning compared with conventional SBs. STATEMENT OF CLINICAL RELEVANCE: These findings support the potential of calibrated SGs to improve digital FA implant accuracy and consistency, aligning with emerging literature on calibrated and framework-based scanning approaches.
Cite this paper
Formatted from the catalogue record. Check names, capitals and page numbers against the paper before you submit.
APA 7
Lerner, H., Weigl, P., Sader, R., Klein, M., Polack, M. A., & Tuminelli, F. J. (2025). Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study. Journal of Dentistry, 164, 106254. https://doi.org/10.1016/j.jdent.2025.106254
MLA 9
Lerner, Henriette, et al. “Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study.” Journal of Dentistry, vol. 164, 2025, p. 106254. https://doi.org/10.1016/j.jdent.2025.106254.
Chicago
Lerner, Henriette, Paul Weigl, Robert Sader, Michael Klein, Mariano A Polack, and Frank J. Tuminelli. 2025. “Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study.” Journal of Dentistry 164: 106254. https://doi.org/10.1016/j.jdent.2025.106254.
Harvard
Lerner, H. et al. (2025) ‘Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study’, Journal of Dentistry, 164, p. 106254. doi:10.1016/j.jdent.2025.106254.
IEEE
H. Lerner, P. Weigl, R. Sader, M. Klein, M. A. Polack, and F. J. Tuminelli, “Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study,” Journal of Dentistry, vol. 164, p. 106254, 2025, doi: 10.1016/j.jdent.2025.106254.
BibTeX
@article{lerner2025accuracy,
title = {{Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study}},
author = {Lerner, Henriette and Weigl, Paul and Sader, Robert and Klein, Michael and Polack, Mariano A and Tuminelli, Frank J.},
journal = {Journal of Dentistry},
year = {2025},
volume = {164},
pages = {106254},
doi = {10.1016/j.jdent.2025.106254},
url = {https://doi.org/10.1016/j.jdent.2025.106254}
}
Download for a reference manager: BibTeX (.bib) · RIS for Zotero, Mendeley and EndNote (.ris)
Topic and keywords
Health Sciences › Dentistry › Oral Surgery › Dental Implant Techniques and Outcomes
- intraoral scanners
- dimensional accuracy
- inter-implant distance
- DESS
- IPD
- ELO
- coordinate measuring machine
Work with this paper
Most-cited papers in Dentistry
- Periodontal Disease in Pregnancy II. Correlation Between Oral Hygiene and Periodontal ConditionJohn Silness, Harald A. Löe · Acta Odontologica Scandinavica · 1964 · 8,071 citations
- Periodontal Disease in Pregnancy I. Prevalence and SeverityHarald A. Löe, John Silness · Acta Odontologica Scandinavica · 1963 · 7,339 citations
- Microbial complexes in subgingival plaqueS. S. Socransky, Anne D. Haffajee, MaryAnn Cugini and 2 more · Journal Of Clinical Periodontology · 1998 · 5,207 citations
- Development of a Classification System for Periodontal Diseases and ConditionsGary C. Armitage · Annals of Periodontology · 1999 · 5,198 citations
- Diagnostic Criteria for Temporomandibular Disorders (DC/TMD) for Clinical and Research Applications: Recommendations of the International RDC/TMD Consortium Network* and Orofacial Pain Special Interest Group†ERIC L. SCHIFFMAN, Richard K. Ohrbach, Edmond L. Truelove and 27 more · Journal of Oral & Facial Pain and Headache · 2014 · 4,942 citations
Questions about this paper
How do I cite this paper in APA style?
Lerner, H., Weigl, P., Sader, R., Klein, M., Polack, M. A., & Tuminelli, F. J. (2025). Accuracy of full-arch implant scan with Nexus IOS® scan gauges versus different conventional scan bodies: An in vitro comparative study. Journal of Dentistry, 164, 106254. https://doi.org/10.1016/j.jdent.2025.106254
How many times has this paper been cited?
9 times, according to OpenAlex on 17 November 2026. Google Scholar and Scopus count different sets of sources, so their numbers differ.
Is this paper free to read?
Yes. It is open access at Journal of Dentistry: sign in and open the full text from this page.
Who wrote this paper?
Henriette Lerner, Paul Weigl, Robert Sader, Michael Klein, Mariano A Polack, Frank J. Tuminelli.
What is the DOI of this paper?
10.1016/j.jdent.2025.106254, permanently at https://doi.org/10.1016/j.jdent.2025.106254. Sign in and open it from this page.
Bibliographic data from OpenAlex (CC0), last checked 17 November 2026. ResearchGuru does not host this paper; once you are signed in, the links open it at the publisher or the repository that does.